In the European Union, Regulation (EU) 2017/746 (IVDR) governs in vitro diagnostic devices and Regulation (EU) 2017/745 (MDR) governs medical devices. This guide summarizes what they mean for a company that develops an instrument and outsources its manufacturing. It does not replace advice from a regulatory specialist or your notified body.
IVDR or MDR: which one applies?
- IVDR: devices intended for the in vitro examination of human specimens (analyzers, reagents, related software, etc.).
- MDR: medical devices in the broad sense, including much equipment used on or for the patient.
Qualification and risk class depend on the device’s intended purpose: that is the first question to settle.
Key dates
- The MDR has applied since 26 May 2021 and the IVDR since 26 May 2022.
- Devices already covered by the former directives can benefit from extended transition periods, under conditions: until the end of 2027 or 2028 depending on class under the MDR (Regulation 2023/607), and until the end of 2027, 2028 or 2029 depending on class under the IVDR (Regulation 2024/1860).
- These extensions require, among other things, a compliant quality system, an application lodged with a notified body and no significant change to the device.
What it means for manufacturing
- Traceability: components, lots and serial numbers must be tracked.
- Manufacturing documentation: instructions, tests and records must be complete and current.
- Supplier control: the legal manufacturer must control critical subcontractors, who may be audited.
- Change management: any design or process change must be assessed.
The role of a manufacturing partner
MGA MedTech does not issue CE marking and does not replace your notified body. As a contract manufacturer, we provide documented, traceable manufacturing: digital manufacturing record, unit testing, recorded results and a shared platform for critical components and documents useful for your file. Our contract manufacturing.
Frequently asked questions
Who is responsible for conformity?
The legal manufacturer, the company placing the device on the market under its name, even when manufacturing is subcontracted.
Does every lab instrument fall under the IVDR?
No. It depends on intended purpose: an instrument specifically intended for in vitro diagnostic use falls under it; a general laboratory instrument may not.
Explore our services
- Medical device CDMO
- Contract manufacturing
- Product development and design
- Engineering services
- Industrialization and design transfer
- Manufacturing transfer
- Prototyping
- Pilot production
- Test benches
- Medical device automation
- IVD instruments
- Laboratory instrument development and manufacturing
- Custom equipment for the pharmaceutical industry
- Medtech startups
- Case study: Nio+ digital PCR with Stilla
- Our Industry 4.0 factory
- How to choose a medtech CDMO
- Guide: medical device outsourcing
Preparing to manufacture an IVD or medical device? Contact our team.